LOGGER-8-RS232: 

RS232 Soil Moisture Sensor Reader - 8 Inputs

This low-cost no-frills data logger is perfect for reading up to 8 Vegetronix soil moisture sensors.  It reports sensor data back to a host computer via RS232, or it can run stand alone and store data to non-volatile memory, for later retrieval. 

LOGGER-8-USB 8 channel soil moisture sensor probe data logger.
LOGGER-8-RS232: 8 Input Data Logger

Setup is very quick, with screwless terminal blocks. It can powered with a wide supply range of 5 to 24V from batteries or a wall transformer.  

Each sensor channel uses 3 spots on the terminal block: for power, ground and the sensor input.  The logger will turn on power to all sensors simultaneously with a programmable setup time.  

The device looks like a serial port to host PC systems so any communications software program such as Termite can be used to control it and gather data.

Data is returned in comma separated format, for easy manipulation in spread sheet programs. Here's an example output file:  LoggerFormat.txt. You can cut and paste data received by Termite into a text file.  You can specify unit labels for each channel, and each channel can be calibrated with a 2 to 10 point calibration curve for non-linear devices. 

The logger can be used with sensors other than our soil moisture probes, so long as the sensors output less than 3.3V.  For sensors with a higher output voltage, a simple resistor divider can be used to scale the voltage to appropriate levels for the logger. 

The logger typically consumes only 5.1mW when it is not sampling data, so it is perfect for battery powered applications. 

Applications

  • Sprinkler and irrigation systems. 
  • Green house irrigation systems.
  • Home automation systems. 
  • General purpose data logging. 

Features

  • Monitors up to 8 sensors.
  • Streams data to a PC, or saves to internal non-volatile memory.  
  • Real time clock. 
  • Low power for battery powered operation.
  • Low cost.
  • Compact.
  • Screwless terminal blocks. 
  • Flanged box for easy wall mounting. 
  • Upgradeable firmware through RS232 port.
  • Each channel can be calibrated with up to a 10 point curve. 

Other Vegetronix Products of Interest

Pricing and Ordering Info

For volume pricing contact us.

ORDER INFO 
Part Number Description  Price Purchase
LOGGER-8-RS232 RS232 Soil Moisture Sensor Data Logger with 8 inputs. $244.95 Buy LOGGER-8-RS232 Now

Specifications

LOGGER-8-RS232

Power consumption 5.1mW @ 6V
Sample Bits 10
Supply voltage 5 to 24V supply
Dimensions 5.0"x2.5"x1.0"   (12.7cm x 6.35cm x 2.54cm)
Number of sensor interfaces 8
Maximum sensor input voltage. 0 to 3.3V
Voltage to Sensors Same as voltage input to unit. 
Memory 91,900 8 channel records.
baud rate 115200
Temperature Range -40°C to 85°C
Enclosure Wall mounting bracket. Not waterproof.

Software Setup

A serial communications program must be used to access the logger.  While the logger will work with most any serial communications program, we recommend Termite for Windows and only provide support for it.  If you are having dificulties, get it working with termite first, then switch to your desired program. 

When you plug a USB Comm device into your computer it will randomly assign a comm port number.   To determine which comm port the device has been assigned to do the following:

  1. Open Windows device manager.  This can be access through the control pannel.
  2. Plug in the USB device.  You should see the device under the tree node labled: "Ports (COM & LPT)"
  3. When you un-plug the USB device, that device should disapear. You now know the correct comm number.
Now that you have the correct comm port number for the device, you can setup Termite:
  1. Open the Termite Program. 
  2. Push the "Setting" button on the too bar.
  3. Select the comm port that you determined from the steps above.
  4. Set the Baud Rate to 9600. 
  5. Data bits should be 8.
  6. Stop bits should be 1.
  7. Parity should be none.
  8. Flow Control should be none.
  9. Under the "Transmitted Text" group box, select "Append CR".
  10. Select Okay to close the Settings box.
  11. Put the cursor on transmit text line, at the very bottom, and type random text, and hit the enter key.  The logger should respond with an error message.  At this point you know that the settings are correct, and you can start sending correctly formatted commands.

Commands

Data in flash memory is stored as a circular queue.  When the memory is full, newer data will overwrite older records.

When the logger powers on. It may take up to 20 seconds for it to initialize the flash memory.  

The default sample rate is once per minute. Power is applied across all sensor channels, and they are sampled at the same instant. 

h<CR> Displays the help menu. 
v<CR> Displays the current software version.
o<CR> Toggle the streaming of data output, on or off.  The default is to stream data out. 
d<CR> Returns the current date. 
d [DD/MM/YYYY]<CR> Sets the current date. 
Example:
d 25/12/2013
t<CR>  Returns the current time.
t [HH:MM:SS]<CR>  Sets the current time. 
Example:
t 13:21:00
b<CR> Enable or disable the heartbeat LED. The purpose of the heart beat LED is to let you know that the board is running.  The reason you would turn off the LED would be to conserve battery power. 
u [sensor index] 'string[5]'<CR> Sets the units of measurement for designated probe. Probes are numbered 1 to 8.  Single quotes are necessary around the string.  Use an empty string to remove units. For example ''.

Example for unit of volts:
u 1 'V'
p [secs]<CR>  Sets or returns the sensor power on time before sampling in seconds.  This allows power to stabilize before sampling the sensors.  All sensors are powered together.

Example of 10 seconds:
p 10
c [sensor index]<CR> Returns the probe calibration table for the specified sensor.
c [sensor index] clear<CR> Clear the probe calibration table for the specified sensor. When no table is used the raw voltage sensed is returned.  
c [sensor index] (X,Y)<CR> Sets the probe calibration table. X is sensor voltage, Y is mapped value. From 2 to 10 points may be entered into the calibration table for each sensor. To enter multiple points just reinvoke this command for each point for the selected sensor. The logger will automatically sort the points by voltage. 
s<CR> Gets the sample period in minutes and seconds.
s [MM:SS]<CR> Sets the sample period in minutes and seconds. Use 00:00 to turn off sampling.

Example of 10 second sample period:
s 0:10
e [Sensor index]<CR>              Toggles enable/disable for a designated sensor.
get<CR> Retrieve all valid data starting with the oldest valid record and ending with the newest record.  We use a slow baud rate to conserve power.  Retrieving all data in the internal flash memory, may take up to an hour. Use a "q<CR>" to terminate this command.
dump<CR> Dump all valid records stored in flash, in non-sequential order. This command normally is not invoked. Because of the slow baud rate, this command may take up to 2 hours.  The only way to terminate the command before it completes is to remove power from the logger. Use a "q<CR>" to terminate this command.  
del<CR> Delete all data in  the internal memory.
wipe<CR> Erase all flash memory.  Use this to purge old data.

Calibration Table and Units.

One of the cool features of the reader is how each channel can be programmed with up to a 10 point calibration table, so that the sensor can be linearized. If no calibration table is entered for a given sensor channel the reader returns the voltage for the channel.  

If your sensor is linear, then only 2 calibration points are required.  For example if 0.0V maps to 0, and 3V maps to 1.5 you would set up the calibration using the following command for sensor 2:

c 2 (0,0)<CR>

c 2 (1.5,3.0)<CR>

In addition, you can specify the units for each channel as a string. 

Wiring Table

Terminal Block TB1 (Sensor Inputs 1 and 2)

Pin Description
1 Sensor 1: Power 
2 Sensor 1: Input
3 Sensor 1: Gnd
4 Sensor 2: Power 
5 Sensor 2: Input
6 Sensor 2: Gnd

Terminal Block TB2 (Sensor Inputs 3 and 4)

Pin Description
1 Sensor 3: Power 
2 Sensor 3: Input
3 Sensor 3: Gnd
4 Sensor 4: Power 
5 Sensor 4: Input
6 Sensor 4: Gnd

Terminal Block TB3 (Sensor Inputs 5 and 6)

Pin Description
1 Sensor 5: Power 
2 Sensor 5: Input
3 Sensor 5: Gnd
4 Sensor 6: Power 
5 Sensor 6: Input
6 Sensor 6: Gnd

Terminal Block TB4 (Sensor Inputs 7 and 8)

Pin Description
1 Sensor 7: Power 
2 Sensor 7: Input
3 Sensor 7: Gnd
4 Sensor 8: Power 
5 Sensor 8: Input
6 Sensor 8: Gnd

Terminal Block TB5 (Power and Commons)

Pin Description
1

Vin (5 to 24VDC)

2 GND


The following picture shows how to hook the logger to a VH400 soil moisture sensor.

Data Logger-8 Hookup
LOGGER-8 Connected to a VH400 Soil Moisture Sensor
(Sensor sold separately)

Firmware Upgrades via RS232 Port

The reader's firmware may be easily upgraded as new firmware versions become available.  To upgrade the reader. Open up the reader box, and set the 2 jumpers on the board.  Download the  latest firmware and the VegeUpdater.exe program.  Connect the reader to the RS232 port, and run the VegeUpdater.exe program in Windows.  Remove the jumpers, and close up the box. 

Trouble Shooting

Problem Solution
The battery is wearing down after a couple of days. Turn off the blinking LED.
The logger doesn't responding to commands. Make sure you are only sending a Carriage Return <CR>, and not also a Line Feed <LF>.  Many comm programs will automatically append a LF to every CR.  Make sure this option is turnned off.
The logger doesn't responding to commands. Make sure you have selected the correct Comm Port. Open windows device manager, and make sure that the logger is listed under the ports section of the device tree.
Random characters or giberish are displayed on the screen. Make sure you have the right baud rate selected.

Wireless Options

The logger will work with any point to point RS232 data radio/modem for remote monitoring.  Digi Corp (was Maxstream) offers several suitable data radios.  For example their XCite™ RF Modems work as an RF link to the RS232 Logger.  Contact Digi Corp for details on their radios and their tech support. 

Battery Life Calculations

Batery life depends on numerous factors including, sample rate, number of sensors, sensor power on time, sensor power, and the sleeping power of the logger.  Depending on how much power is used each day you can compute the capacity of your batteries.  Batteries are rated in mA-hours which is roughly speaking the number of hours the battery will run when disipated by 1mA.  To calculate the proper battery size, and the number days a battery will last you can use an online calculator to estimate it: Battery-Life-Calculator.

It's always wise to leave extra margin by using a larger battery capacity than estimated.

How to Get Started

The best way to get started is to order a reader board and a few low cost soil probes and try them out in your application. 